Safety: CONDITIONAL
Bottom line: MRI permissible within designated field strengths. Requires radiologist approval and scheduling.
Leadless Cardiac Pacemaker
Safety: Conditional
Risks: Thermal, Arrhythmia, Device Malfunction
Conditions:
- Identify type of leadless pacemaker to determine maximum static field strength
- Medtronic Micra Pacemaker
- Static field strength up to and including 3T
- St. Jude Nanostim Leadless Pacemaker
- Static field strength up to and including 1.5T
- Medtronic Micra Pacemaker
- Requires cardiologist programming before and after MRI
- Do not bring controller into MRI scanner
Comments:
- Also known as Micra pacemaker or Nanostim pacemaker
- Not to be confused with Implanted Cardiac Event Monitor (Loop Recorder), which may be mistaken for Leadless Pacemaker on radiograph
- For scheduling of cases, please see Scheduling for Patients with MRI Conditional Devices.
- Currently these devices are only imaged on the Montefiore Moses campus
See Also:
Bottom line: MRI permissible within designated field strengths. Requires radiologist approval and scheduling.