Safety: CONDITIONAL
Bottom line: MRI permissible if >6 weeks, or <6 if weeks if conditions are met
Stent Grafts
Safety: Conditional
Risks: Thermal
Conditions:
- If implanted >6 weeks, it is MRI permissible
- If implanted <6 weeks, all manufacturer’s conditions must be met to be MRI compatible
Comments:
- MR conditional stent grafts may be scanned immediately after implant, if in accord with manufacturer’s approval
- Patients with unknown stent grafts should wait 6 weeks
- Because of fluid flow through the stent, there is minimal risk of heating
Bottom line: MRI permissible if >6 weeks, or <6 if weeks if conditions are met