Safety:  UNSAFE

Bottom line:  Do NOT perform MRI

Abandoned Intravascular Pacer and ICD Leads

Safety:  Unsafe

Risks:  Thermal

Conditions:  None

Comments:

  • No pulse generator attached, ONLY abandoned intravascular leads. Abandoned intravascular pacer and ICD leads, even if from MRI conditional pacing systems, are MRI unsafe.
  • Intravascular lead fragments which do not have an intracardiac or proximal great vessel component are permissible, such as axillary vein fragments retained following cardiac transplant.

See Also:

Bottom line: Do NOT perform MRI. Exception for retained intravascular leads without an intracardiac component.